World-Class Optical Performance
- Wavelength range: 600 to 1700nm
- High wavelength accuracy: ±0.008 nm typ.
- High wavelength resolution: 0.02nm
- Wide dynamic range: 78dB typ.
- Wide level range: +20 to -90dBm
- Fast measurement: 0.2 sec. (100nm span)
- Applicable to single-mode and multimode fibers
| Product Shipping |
|
Product Description
The AQ6370E OSA has a versatile wavelength range of 600 nm to 1700 nm, ideal for both telecom and general purpose applications. The unique free-space input design allows testing of both DWDM-class single-mode and VCSEL-sourced multimode fibers in a single model and the high-speed measurement mode, high close-in dynamic range mode, touch panel, and APP function greatly improve measurement efficiency.

Standard and High-Performance Models
There are two models available, Standard and High performance. The High performance model provides even higher wavelength accuracy and dynamic range.
High wavelength resolution: 0.02nm
High Wavelength Accuracy: ±0.008 nm typ.
- High performance model: ±0.008 nm typ. (C band)
- Standard model: ±0.015 nm typ.(S+C+L band)
| Wavelength range | Standard (-10) | High performance (-20) |
|---|---|---|
| 1520 to 1580 nm | ±0.015 nm | ±0.008 nm |
| 1450 to 1520 nm 1580 to 1620 nm | ±0.015 nm |
Note: The wavelength accuracy values in the table are typical values.
Ultra-High Dynamic Range: 78dB typ.
With the reduced stray-light in the monochromator, AQ6370E achieves ultra-high dynamic range of typ. 78dB.
| Standard (-10) | High Performance (-20) | |
| Peak± 1.0 nm Peak± 0.4 nm Peak± 0.2 nm | 73 dB 62 dB 45 dB | 73 dB (Typ.78dB) 64 dB (Typ.70dB) 50 dB (Typ.55dB) |
| *Resolution setting 0.05 nm | ||
Sharper spectrum measurement: HCDR mode
The HCDR (High Close-in Dynamic Range) mode is a feature for single longitudinal mode laser measurements that makes the spectrum around the peak sharper and the side modes more clearly visualized.
This mode is only available on the High performance model(-20).